ELECTRICAL SAFETY ANALYZER SA-2010 SA-2010-INTL SA-2010-AUS USER MANUAL
2 AVAILABLE MODELS: SA-2010 – STANDARD MODEL WITH HOSPITAL-GRADE NEMA 5-15P LINE PLUG AND HOSPITAL-GRADE NEMA 5-20R DUT TEST RECEPTACLE FOR USE
3 BC20-20200 REPLACEMENT SA-2010-INTL DUT INTERNATIONAL RECEPTACLE ADAPTER BC20-20221 SCHUKO GROUNDING ADAPTER BC20-41337 RS-232
4 This section looks at the layouts of the SA-2010 Models and gives descriptions of the elements that are present. For all models, the user contro
5 SA-2010-INTL Layout SA-2010-AUS Layout (SA-2010-INTL-R shown)
6 Display The main information in the system is presented in the 3-1/2 digit LCD display. This data is provided as simple meter readings with the
7 Power Outlet Indicators Three Neon indicators help verify the polarity and wiring of the wall receptacle that the Safety Analyzer is plugged into
8 Communications (Optional) On Analyzer models with the –R option, a DB-9 RS-232 Communications
9 The SA-2010 Models allow the user a great deal of flexibility in testing. Any of the basic tests can be run and in almost any sequence. The inf
10 Device Current With the Device Current function selected, the display will show the current draw of the DUT. The Receptacle should be configured
11 Lead to Earth/Ground Leakage With the Lead to Earth/Ground function selected, the display will show the leakage current between the
12 As each lead and then All leads are selected, depress and hold the “ISO” Test key. This will apply 110% of the line voltage through
13 The test requires that the Chassis Test lead be plugged into the Chassis Connector and the External Test lead be plugged into the External Connec
14 The communication protocol provides a means to completely configure and use the Analyzer from a PC or other device with RS-232 communications in
15 uppercase and lowercase. Commands sent to the device are not case sensitive. Upper case characters in the command summary table below are used
16 Revision # Revisions Made Rev 01 Origination P:\MANUALS\BCGroup\…\SA2010_Series_UM_Rev01.docx
17 MAINS VOLTAGE RANGE 90 to 264 VAC, 50/60 Hz ACCURACY ± 3% of Reading, ± 1 digit DEVICE CURRENT RANGE 0 to 19.99 Amps, RMS ACCURACY ± 5% of
18 ELECTRICAL OPERATING LINE VOLTAGE 90 to 264 VAC, 50/60 Hz LINE PLUG SA-2010 MODELS NEMA 5-15P SA-2010-INTL MODELS Universal IEC C20 Receptacle P
19 ELECTRICAL (continued) DUT RECEPTACLE (continued) SA-2010-INTL MODELS o EU1-16P (EURO) CEE 7/7 “SCHUKO” (NOTE: MUST USE BC20-20221
20 NOTES
21 NOTES
i WARNINGS, CAUTIONS, NOTICES ... ii DESCRIPTION ...
22 NOTES
BC GROUP INTERNATIONAL, INC. 3081 ELM POINT INDUSTRIAL DRIVE ST. CHARLES, MO 63301 USA 1-800-242-8428 1-314-638-3800 www.bcgroup
ii WARNING - USE The SA-2010 is intended for testing only and should never be used in diagnostics, tre
iii CAUTION - SERVICE The SA-2010 is intended to be serviced only by authorized service personnel. Troubleshooting and service procedures should on
iv CAUTION - FUSE Only replace the SA-2010 fuse with the specified type and rating.
v NOTICE – SYMBOLS Symbol Description Caution (Consult Manual for Further Information) Electrical Caution (Consult Manual for Further
vi SA-2010 Series User Manual Copyright © 2012 www.bcgroupintl.com Made in the USA 08/12
1 The Model SA-2010 Series is a Microprocessor based Electrical Safety Analyzer. It allows for a multitude of tests to be performed on a