MULTI-PARAMETER PATIENT SIMULATOR PS-2200 SERIES USER MANUAL
2 S-T SEGMENT ELEVATION: ± 0, 0.05, 0.1, 0.2, 0.3, 0.4, 0.5, 0.6, 0.7, 0.8 mV ARTIFACTS: 50 Hz, 60 Hz, MUSCLE, BASELINE WANDER, RESPIRAT
3 SQUARE: 0.125, 2.0 Hz TRIANGLE: 2.0, 2.5 Hz PULSE: 30, 60, 120 BPM; 60 ms WIDTH AMPLITUDE (Lead II): 0.05, 0.10, 0
4 A special Swan-Ganz mode is available that will allow Manual or Automatic simulation of that sequence. Each channel has independent settings.
5 SPO2 SIMULATION (Optional) When used with the MSP-2100 external module and FingerSim family of SpO2 finger simulators, the system will provide a
6 OPTIONAL ACCESSORIES BC20-21100 BATTERY ELIMINATOR (USA Version) BC20-21101 BATTERY ELIMINATOR (Euro Version) BC20-02300 FETAL / MATERNAL O
7 This section looks at the layout of a PS-2200 and gives descriptions of the elements that are present. 8 Light
8 General Operation The unit is controlled by 17 light touch keys. They allow the user to move around within the displayed parameters, select the d
9 The key opens a screen that allows the user to select the unit’s general output settings, as well as setup for the system.
10 Power Key The key turns the unit on and off. To turn off the unit, the key must be held for 1 second. Backlight The Graphic LCD dis
11 Respiration Respiration waveforms are provided that have adjustable rates from 15 to 120 BrPM (Breaths Per Minute) as well as an Apnea (0 BrPM) s
12 High Level Output (+) A high level ECG output signal (200 x Amplitude Setting) is available in the BP1 6-Pin mini-DIN connector. A
13 continue use of the unit. If the battery is not replaced before the battery reaches a critical level (0 % Battery Life), the unit will shut down
14 The PS-2200 can send waveforms to ECG machines in 3, 5 or 12 lead configurations. It has independent outputs for each signal lead, referenced t
15 The rate, amplitude, adult/pediatric, artifact and ST elevation or depression can be selected by using to highligh
16 The PS-2200 can send Arrhythmia waveforms to ECG machines in 3, 5 or 12-lead configurations. It has independent outputs for each
17 The top portion of the display shows the currently enabled arrhythmia group and selection, selectable from: Premature Supraventricular Ventric
18 The grouping, arrhythmias and amplitude can be selected by using to highlight the parameter to change and using
19 The following is a brief description of how the PS-2200 simulates the available Arrhythmias: PREMATURE Abbreviation Arrhythmia Descr
20 PVC 2 – Auto Standard Type 2 Premature Ventricular Contraction NSR of 80 BPM with periodic right focus premature ventricular beats with 20% prem
21 SUPRAVENTRICULAR Abbreviation Arrhythmia Description Atrial Fib – Coarse Artial Fibrillation Absence of P-wave, irregular P-R interval rate and
i WARNINGS, CAUTIONS, NOTICES ... ii DESCRIPTION ...
22 VENTRICULAR Abbreviation Arrhythmia Description Pair of PVCs – Auto Pair of Premature Ventricular Contractions NSR of 80 BPM with Periodic Grou
23 Vent Tach Ventricular Tachycardia 160 BPM, No P-wave, Beats similar to Type 1 PVC (Continuous) Vent Fib – Coarse Ventricular Fibrillation Irreg
24 CONDUCTION Abbreviation Arrhythmia Description 1st Deg Heart Block First Degree Heart Block 80 BPM with a long P-R interval of 250 ms (Continuo
25 The PS-2200 can send paced waveforms to ECG machines in 3, 5 or 12 lead configurations. It has independent outputs for each signa
26 The pacemaker rhythms and signals can be selected by using to highlight the parameter to change and using
27 The PS-2200 can send performance waveforms to ECG machines in 3, 5 or 12-lead configurations. It has independent outputs for each signal lead,
28 R-Wave When one of the 6 R-Wave waveforms is selected, the display changes to allow the setting of the width. These widths can be sel
29 Auto Wave If the Performance parameter is set to AUTO, the unit will automatically sequence through all of the performance waves, starting with S
30 NOTE: The Transducer Sensitivity (5 or 40 µV/V/mmHg) must be set to correlate with the monitoring equipment before simulation can begin. (See
31 The PS-2240 quad blood pressure display will resemble the following: These settings can be selected by using
ii WARNING - USERS The PS-2200 Series are for use by skilled technical personnel only. WARNING - USE The PS-2200 Series are intended for testing
32 Alternately, to see a submenu of all the options for a highlighted parameter, use . Use to scroll to the desired
33 Auto Sets If Auto Preset is selected, the channels will automatically sequence through the static sets, starting with Set 0, incrementing at a fi
34 The PS-2220 dual blood pressure display will resemble the following: These settings can be selected by using
35 Alternately, to see a submenu of all the options for a highlighted parameter, use . Use to scroll to the desire
36 The PS-2210 single blood pressure display will resemble the following: or These settings can be selected by us
37 Alternately, to see a submenu of all the options for a highlighted parameter, use . Use to scroll to the desired
38 The Swan-Ganz simulation is a special feature that will run the typical sequence for a Swan-Ganz catheter. This can be done either manually, w
39 Manual In the Swan-Ganz – Man mode, the user controls the sequence using the keys. (The functions of the key
40 Automatic In the Swan-Ganz – Auto mode, the unit continually runs the sequence. The time remaining before proceeding to the next
41 The following is a typical running screen: To pause or exit the Swan-Ganz – Auto Mode during a sequence, use . The fo
iii CAUTION - ENVIRONMENT Exposure to environmental conditions outside the specifications can adversely affect the performance of the PS-2200 Series
42 NOTE: The delta ohm Respiration Signal can be inserted in either the LL or LA lead. The Baseline impedance can be set to 500, 1000, 1500 or 2
43 These Rates and Amplitude (Impedance Variations) can be selected by using to highlight the parameter to change an
44 The PS-2200 simulates 7 temperatures that are independent from the rest of the functions of the unit. The temperature setting can be selected
45 The PS-2200 provides the unique feature of a Training Mode to aid the user in practicing the identification of arrhythmias. The unit will sequ
46 Alternately, to see a submenu of all the options for a highlighted parameter, use . Use to scroll to the desir
47 SUBSET The subset feature allows the user to select specific arrhythmias for a more controlled training. This feature is selected by setting the
48 3. To select all the arrhythmias in a group, use to scroll to the category and to toggle the indicator to “ALL”.
49 5. When completed selecting the desired arrhythmias from that group, use to return to the Group Submenu. “SOME” will appear to indicate
50 The PS-2200 allows for setup of the Outputs and the System Parameters through the category key. Depress the key multiple times to
51 System Setup The System Setup screen allows for the setting of the parameters controlling various function of the unit as well as the viewing of
iv NOTICE – CE The PS-2200 Series Simulators bear the mark Based on the following testing standards: ELECTROMAGNETIC COMPATIBILITY DIRE
52 The following is a brief description of the parameters and the available range of settings: Parameter Description Range Auto Off Timer The elaps
53 The PS-2200 allows the user to tailor the settings that the unit will have on Power Up. The “Power Up With” parameter in the System Setup Menu
54 SystemSetup: Auto Timer Off 30 min Backlight Time 5 sec Backlight Intensity 100% Contrast Adjust 10 Power Up With Default Auto
55 The PS-2200 has the ability to drive an external SpO2 module. This module (MSP-2100) accepts the FingerSim family of SpO2 finger simulators (f
56 The PS-2200 has the ability to simulate a combined Fetal/Maternal ECG. It assumes a Fetal Scalp Electrode and a Maternal Thigh Electrode. Add
57 The display will resemble the following: NOTE: The Transducer Sensitivity (5 mV/V/mmHg or 40 mV/V/mmHg) must be set to co
58 IUP Simulation This feature is designed to simulate the IUP and corresponding FHR that might occur during labor. The base 140 BPM FHR will respo
59 The PS-2200 has the ability to simulate Thermodilution Cardiac Output measurements. Thermodilution allows the calculation of heart volume outpu
60 Setting Up a Test 1) The DUT (Cardiac Output Monitor) needs to be setup to match the PS-2200. The following settings are required
61 Running a Test 1) Select the desired test. 2) Select the injectate temperature (0 °C or 24 °C). NOTE: This is different than the select
v NOTICE – SYMBOLS Symbol Description Caution (Consult Manual for Further Information) Center Negative Per European Council Directive
62 The following are representations of the socket connectors found on the unit. They are viewed as if looking at the socket in the unit, not the
63 Revision # Program # Revisions Made Rev 01 DT7347CA Preliminary Manual Rev 02 DT7347CA Updates to Preliminary Manual Rev 03 DT7347CA
64 ECG SIMULATION RATE NSR 30, 40, 45, 60, 80, 90, 100, 120, 140, 160, 180, 200, 220, 240, 260, 280, 300 BPM PERFORMANCE WAVEFORMS SINE 0.1, 0.5,
65 PACEMAKER WAVEFORMS RATE 75 BPM ACCURACY ± 1% WIDTH 0.1, 0.5, 1.0, 1.5, 2.0 ms ACCURACY ± 5% AMPLITUDE 1, 2, 3, 4, 5, 6, 7, 8, 9, 10 mV ACCURACY
66 TEMPERATURE SIMULATION SELECTION 0, 24, 30, 37, 40 °C (32.0, 75.2, 86.0, 98.6, 104.0 °F) ACCURACY ± 0.1 °C TYPE YSI Series 400 and 700 FETAL / M
67 PHYSICAL & ENVIRONMENTAL DISPLAY Two 128 X 64 Pixel Graphical LCDs, White LED Backlight CONSTRUCTION ENCLOSURE ABS Plastic FACE PLATE Lexan,
68 NOTES
BC GROUP INTERNATIONAL, INC. 3081 ELM POINT INDUSTRIAL DRIVE ST. CHARLES, MO 63301 USA 1-800-242-8428 1-314-638-3800 www.bcgroup
vi PS-2200 Series User Manual Copyright © 2012 www.bcgroupintl.com Made in the USA 10/12
1 The PS-2200 Family is a series of high end Microprocessor based Patient Simulators. Each model provides ECG, Blood Pressure, Respiration and Te