Page 1 - PATIENT SIMULATOR

MULTI-PARAMETER PATIENT SIMULATOR PS-2200 SERIES USER MANUAL

Page 2

2 S-T SEGMENT ELEVATION: ± 0, 0.05, 0.1, 0.2, 0.3, 0.4, 0.5, 0.6, 0.7, 0.8 mV ARTIFACTS: 50 Hz, 60 Hz, MUSCLE, BASELINE WANDER, RESPIRAT

Page 3 - TABLE OF CONTENTS

3 SQUARE: 0.125, 2.0 Hz TRIANGLE: 2.0, 2.5 Hz PULSE: 30, 60, 120 BPM; 60 ms WIDTH AMPLITUDE (Lead II): 0.05, 0.10, 0

Page 4

4 A special Swan-Ganz mode is available that will allow Manual or Automatic simulation of that sequence. Each channel has independent settings.

Page 5 - CAUTION - LIQUIDS

5 SPO2 SIMULATION (Optional) When used with the MSP-2100 external module and FingerSim family of SpO2 finger simulators, the system will provide a

Page 6 - NOTICE – CE

6 OPTIONAL ACCESSORIES BC20-21100 BATTERY ELIMINATOR (USA Version) BC20-21101 BATTERY ELIMINATOR (Euro Version) BC20-02300 FETAL / MATERNAL O

Page 7 - NOTICE – ABBREVIATIONS

7 This section looks at the layout of a PS-2200 and gives descriptions of the elements that are present. 8 Light

Page 8 - NOTICE – CONTACT INFORMATION

8 General Operation The unit is controlled by 17 light touch keys. They allow the user to move around within the displayed parameters, select the d

Page 9

9 The key opens a screen that allows the user to select the unit’s general output settings, as well as setup for the system.

Page 10

10 Power Key The key turns the unit on and off. To turn off the unit, the key must be held for 1 second. Backlight The Graphic LCD dis

Page 11

11 Respiration Respiration waveforms are provided that have adjustable rates from 15 to 120 BrPM (Breaths Per Minute) as well as an Apnea (0 BrPM) s

Page 13

12 High Level Output (+) A high level ECG output signal (200 x Amplitude Setting) is available in the BP1 6-Pin mini-DIN connector. A

Page 14

13 continue use of the unit. If the battery is not replaced before the battery reaches a critical level (0 % Battery Life), the unit will shut down

Page 15

14 The PS-2200 can send waveforms to ECG machines in 3, 5 or 12 lead configurations. It has independent outputs for each signal lead, referenced t

Page 16

15 The rate, amplitude, adult/pediatric, artifact and ST elevation or depression can be selected by using to highligh

Page 17

16 The PS-2200 can send Arrhythmia waveforms to ECG machines in 3, 5 or 12-lead configurations. It has independent outputs for each

Page 18

17 The top portion of the display shows the currently enabled arrhythmia group and selection, selectable from: Premature Supraventricular Ventric

Page 19

18 The grouping, arrhythmias and amplitude can be selected by using to highlight the parameter to change and using

Page 20 - Description

19 The following is a brief description of how the PS-2200 simulates the available Arrhythmias: PREMATURE Abbreviation Arrhythmia Descr

Page 21

20 PVC 2 – Auto Standard Type 2 Premature Ventricular Contraction NSR of 80 BPM with periodic right focus premature ventricular beats with 20% prem

Page 22 - ECG – NORMAL SINUS RHYTHM

21 SUPRAVENTRICULAR Abbreviation Arrhythmia Description Atrial Fib – Coarse Artial Fibrillation Absence of P-wave, irregular P-R interval rate and

Page 23 - Displays time (seconds)

i WARNINGS, CAUTIONS, NOTICES ... ii DESCRIPTION ...

Page 24 - ECG – ARRHYTHMIAS

22 VENTRICULAR Abbreviation Arrhythmia Description Pair of PVCs – Auto Pair of Premature Ventricular Contractions NSR of 80 BPM with Periodic Grou

Page 25

23 Vent Tach Ventricular Tachycardia 160 BPM, No P-wave, Beats similar to Type 1 PVC (Continuous) Vent Fib – Coarse Ventricular Fibrillation Irreg

Page 26

24 CONDUCTION Abbreviation Arrhythmia Description 1st Deg Heart Block First Degree Heart Block 80 BPM with a long P-R interval of 250 ms (Continuo

Page 27 - PREMATURE

25 The PS-2200 can send paced waveforms to ECG machines in 3, 5 or 12 lead configurations. It has independent outputs for each signa

Page 28

26 The pacemaker rhythms and signals can be selected by using to highlight the parameter to change and using

Page 29 - SUPRAVENTRICULAR

27 The PS-2200 can send performance waveforms to ECG machines in 3, 5 or 12-lead configurations. It has independent outputs for each signal lead,

Page 30 - VENTRICULAR

28 R-Wave When one of the 6 R-Wave waveforms is selected, the display changes to allow the setting of the width. These widths can be sel

Page 31

29 Auto Wave If the Performance parameter is set to AUTO, the unit will automatically sequence through all of the performance waves, starting with S

Page 32 - CONDUCTION

30 NOTE: The Transducer Sensitivity (5 or 40 µV/V/mmHg) must be set to correlate with the monitoring equipment before simulation can begin. (See

Page 33 - ECG – PACEMAKER

31 The PS-2240 quad blood pressure display will resemble the following: These settings can be selected by using

Page 34 - Waveform

ii WARNING - USERS The PS-2200 Series are for use by skilled technical personnel only. WARNING - USE The PS-2200 Series are intended for testing

Page 35 - ECG – PERFORMANCE

32 Alternately, to see a submenu of all the options for a highlighted parameter, use . Use to scroll to the desired

Page 36

33 Auto Sets If Auto Preset is selected, the channels will automatically sequence through the static sets, starting with Set 0, incrementing at a fi

Page 37

34 The PS-2220 dual blood pressure display will resemble the following: These settings can be selected by using

Page 38 - BLOOD PRESSURE

35 Alternately, to see a submenu of all the options for a highlighted parameter, use . Use to scroll to the desire

Page 39 - PS-2240

36 The PS-2210 single blood pressure display will resemble the following: or These settings can be selected by us

Page 40

37 Alternately, to see a submenu of all the options for a highlighted parameter, use . Use to scroll to the desired

Page 41

38 The Swan-Ganz simulation is a special feature that will run the typical sequence for a Swan-Ganz catheter. This can be done either manually, w

Page 42 - PS-2220

39 Manual In the Swan-Ganz – Man mode, the user controls the sequence using the keys. (The functions of the key

Page 43

40 Automatic In the Swan-Ganz – Auto mode, the unit continually runs the sequence. The time remaining before proceeding to the next

Page 44 - PS-2210

41 The following is a typical running screen: To pause or exit the Swan-Ganz – Auto Mode during a sequence, use . The fo

Page 45

iii CAUTION - ENVIRONMENT Exposure to environmental conditions outside the specifications can adversely affect the performance of the PS-2200 Series

Page 46 - SWAN-GANZ

42 NOTE: The delta ohm Respiration Signal can be inserted in either the LL or LA lead. The Baseline impedance can be set to 500, 1000, 1500 or 2

Page 47

43 These Rates and Amplitude (Impedance Variations) can be selected by using to highlight the parameter to change an

Page 48 - (Dependant

44 The PS-2200 simulates 7 temperatures that are independent from the rest of the functions of the unit. The temperature setting can be selected

Page 49 - SEQUENCE

45 The PS-2200 provides the unique feature of a Training Mode to aid the user in practicing the identification of arrhythmias. The unit will sequ

Page 50 - RESPIRATION

46 Alternately, to see a submenu of all the options for a highlighted parameter, use . Use to scroll to the desir

Page 51 - ECG/RESPIRATION/TEMPERATURE

47 SUBSET The subset feature allows the user to select specific arrhythmias for a more controlled training. This feature is selected by setting the

Page 52 - TEMPERATURE

48 3. To select all the arrhythmias in a group, use to scroll to the category and to toggle the indicator to “ALL”.

Page 53 - TRAINING

49 5. When completed selecting the desired arrhythmias from that group, use to return to the Group Submenu. “SOME” will appear to indicate

Page 54 - AMPLITUDE

50 The PS-2200 allows for setup of the Outputs and the System Parameters through the category key. Depress the key multiple times to

Page 55

51 System Setup The System Setup screen allows for the setting of the parameters controlling various function of the unit as well as the viewing of

Page 56

iv NOTICE – CE The PS-2200 Series Simulators bear the mark Based on the following testing standards: ELECTROMAGNETIC COMPATIBILITY DIRE

Page 57

52 The following is a brief description of the parameters and the available range of settings: Parameter Description Range Auto Off Timer The elaps

Page 58 - (See Power Up Settings)

53 The PS-2200 allows the user to tailor the settings that the unit will have on Power Up. The “Power Up With” parameter in the System Setup Menu

Page 59

54 SystemSetup: Auto Timer Off 30 min Backlight Time 5 sec Backlight Intensity 100% Contrast Adjust 10 Power Up With Default Auto

Page 60

55 The PS-2200 has the ability to drive an external SpO2 module. This module (MSP-2100) accepts the FingerSim family of SpO2 finger simulators (f

Page 61 - POWER UP SETTINGS

56 The PS-2200 has the ability to simulate a combined Fetal/Maternal ECG. It assumes a Fetal Scalp Electrode and a Maternal Thigh Electrode. Add

Page 62

57 The display will resemble the following: NOTE: The Transducer Sensitivity (5 mV/V/mmHg or 40 mV/V/mmHg) must be set to co

Page 63 - (Option)

58 IUP Simulation This feature is designed to simulate the IUP and corresponding FHR that might occur during labor. The base 140 BPM FHR will respo

Page 64 - FETAL / MATERNAL (Option)

59 The PS-2200 has the ability to simulate Thermodilution Cardiac Output measurements. Thermodilution allows the calculation of heart volume outpu

Page 65

60 Setting Up a Test 1) The DUT (Cardiac Output Monitor) needs to be setup to match the PS-2200. The following settings are required

Page 66

61 Running a Test 1) Select the desired test. 2) Select the injectate temperature (0 °C or 24 °C). NOTE: This is different than the select

Page 67 - CARDIAC OUTPUT (Option)

v NOTICE – SYMBOLS Symbol Description Caution (Consult Manual for Further Information) Center Negative Per European Council Directive

Page 68

62 The following are representations of the socket connectors found on the unit. They are viewed as if looking at the socket in the unit, not the

Page 69

63 Revision # Program # Revisions Made Rev 01 DT7347CA Preliminary Manual Rev 02 DT7347CA Updates to Preliminary Manual Rev 03 DT7347CA

Page 70 - OUTPUT CONNECTIONS

64 ECG SIMULATION RATE NSR 30, 40, 45, 60, 80, 90, 100, 120, 140, 160, 180, 200, 220, 240, 260, 280, 300 BPM PERFORMANCE WAVEFORMS SINE 0.1, 0.5,

Page 71 - MANUAL REVISIONS

65 PACEMAKER WAVEFORMS RATE 75 BPM ACCURACY ± 1% WIDTH 0.1, 0.5, 1.0, 1.5, 2.0 ms ACCURACY ± 5% AMPLITUDE 1, 2, 3, 4, 5, 6, 7, 8, 9, 10 mV ACCURACY

Page 72 - SPECIFICATIONS

66 TEMPERATURE SIMULATION SELECTION 0, 24, 30, 37, 40 °C (32.0, 75.2, 86.0, 98.6, 104.0 °F) ACCURACY ± 0.1 °C TYPE YSI Series 400 and 700 FETAL / M

Page 73

67 PHYSICAL & ENVIRONMENTAL DISPLAY Two 128 X 64 Pixel Graphical LCDs, White LED Backlight CONSTRUCTION ENCLOSURE ABS Plastic FACE PLATE Lexan,

Page 74

68 NOTES

Page 76

BC GROUP INTERNATIONAL, INC. 3081 ELM POINT INDUSTRIAL DRIVE ST. CHARLES, MO 63301 USA 1-800-242-8428 1-314-638-3800 www.bcgroup

Page 77

vi PS-2200 Series User Manual Copyright © 2012 www.bcgroupintl.com Made in the USA 10/12

Page 78

1 The PS-2200 Family is a series of high end Microprocessor based Patient Simulators. Each model provides ECG, Blood Pressure, Respiration and Te