BC Biomedical NIBP-1020 User Manual

Browse online or download User Manual for Video converters BC Biomedical NIBP-1020. BC Biomedical NIBP-1020 User manual

  • Download
  • Add to my manuals
  • Print

Summary of Contents

Page 1 - Simulator

Non-Invasive Blood Pressure Simulator NIBP-1010 NIBP-1020 USER MANUAL

Page 2

8 MAIN SCREENS – There are five main screens: Pressure Only, Pressure with Output Waveform, Pressure with Pressure Graph, ECG (optional) and Batter

Page 3 - TABLE OF CONTENTS

9OUTPUT WAVEFORM– This screen shows the output waveform as well as the pressure, peak pressure, and output selection. The display will resembl

Page 4 - WARNING

10 PRESSURE GRAPH SCREEN – This screen provides a graph of the pressure, as well as, the pressure, peak pressure, and output selection PRES

Page 5 - BC GROUP

11ECG OUTPUT SCREEN – This screen shows the selected ECG output mode. (NIBP-1020 units only) NOTE: While in this mode, the NIBP simulation does

Page 6

12BATTERY INDICATOR SCREEN – This screen shows the status of the battery voltage level with Built-In Battery Option only. NOTE: It is only an esti

Page 7 - OVERVIEW

13 The Setup Mode allows the user to adjust the configuration of the simulator. The Setup screen can be entered using the key. The param

Page 8

14SYSTOLIC AND DIASTOLIC SHIFT – The NIBP-1000 Series is equipped with the option to shift test results to compensate for different methods of measu

Page 9

15 The NIBP-1010 and NIBP-1020 Blood Pressure Simulators are connected between the blood pressure cuff and the monitor (Device Under Test, DUT).

Page 10 - SCREENS

16 For the NIBP-1020, with ECG output, test snaps are provided along the sides and are identified by the markings on the overlay.

Page 11 - (Alternates every 30 sec)

17 A motor and piston assembly is used to generate the output waveforms. A differential pressure sensor is used to measure the pressure of the cuf

Page 13 - SELECTED

18 NIBP: To run an NIBP simulation, the cuff and monitor are connected to the pressure input. Then the measurement is initiated by the monitor an

Page 14 - CAUTION

19 Revision # Program # Revisions Made Rev 01 DT7355CA Origination Rev 02 DT7355CA Specification Information Updated Rev 03 DT7355CA Miscell

Page 15

20 NIBP-1010 Blood Pressure RANGE 0-500 mmHg @ 20°C ACCURACY +/- (1% of Reading + .5 mmHg) RATE 80, 94

Page 16

21NIBP-1020 Blood Pressure RANGE 0-500 mmHg @ 20°C ACCURACY +/- (1% of Reading + .5 mmHg)

Page 17 - OPERATIONS

22 P:\Manuals\BCGroup\…\NIBP1000_UM_Rev06.doc Respiration RATE ACCURACY +/- 1% IMPEDANCE DE

Page 20 - RUNNING A TEST

1 DESCRIPTION... 3 OVERVIEW...

Page 21 - LIMITED WARRANTY

2 Manual NIBP-1000 Series Copyright © 2006 www.bcgroupintl.com Made in the USA 7/06

Page 22 - SPECIFICATIONS

3 The Model NIBP-1000 Series is a family of Microprocessor based, High Precision Non-Invasive Blood Pressure (NIBP) Simulators. The units are smal

Page 23 - Pacemaker Waveforms

4AVAILABLE MODELS: NIBP1010 BASIC UNIT NIBP1010-P + PEAK NIBP1010-BP + BATTERY & PEAK NIBP1020 BASIC UNIT WITH EKG NIBP1020-PA

Page 24 - Electrical

5 This section looks at the layout of the NIBP-1010 and gives descriptions of the elements that are present. OVERVIEW 7 pin Mini-Din connector fo

Page 25

6This section looks at the layout of the NIBP-1020 and gives descriptions of the elements that are present. 7 pin Mini-Din connector for RS-232 and

Page 26

7 Six tactile-touch keys are provided for system operation: – This key turns the unit off and on. The unit will return to the screen that was a