Non-Invasive Blood Pressure Simulator NIBP-1010 NIBP-1020 USER MANUAL
8 MAIN SCREENS – There are five main screens: Pressure Only, Pressure with Output Waveform, Pressure with Pressure Graph, ECG (optional) and Batter
9OUTPUT WAVEFORM– This screen shows the output waveform as well as the pressure, peak pressure, and output selection. The display will resembl
10 PRESSURE GRAPH SCREEN – This screen provides a graph of the pressure, as well as, the pressure, peak pressure, and output selection PRES
11ECG OUTPUT SCREEN – This screen shows the selected ECG output mode. (NIBP-1020 units only) NOTE: While in this mode, the NIBP simulation does
12BATTERY INDICATOR SCREEN – This screen shows the status of the battery voltage level with Built-In Battery Option only. NOTE: It is only an esti
13 The Setup Mode allows the user to adjust the configuration of the simulator. The Setup screen can be entered using the key. The param
14SYSTOLIC AND DIASTOLIC SHIFT – The NIBP-1000 Series is equipped with the option to shift test results to compensate for different methods of measu
15 The NIBP-1010 and NIBP-1020 Blood Pressure Simulators are connected between the blood pressure cuff and the monitor (Device Under Test, DUT).
16 For the NIBP-1020, with ECG output, test snaps are provided along the sides and are identified by the markings on the overlay.
17 A motor and piston assembly is used to generate the output waveforms. A differential pressure sensor is used to measure the pressure of the cuf
18 NIBP: To run an NIBP simulation, the cuff and monitor are connected to the pressure input. Then the measurement is initiated by the monitor an
19 Revision # Program # Revisions Made Rev 01 DT7355CA Origination Rev 02 DT7355CA Specification Information Updated Rev 03 DT7355CA Miscell
20 NIBP-1010 Blood Pressure RANGE 0-500 mmHg @ 20°C ACCURACY +/- (1% of Reading + .5 mmHg) RATE 80, 94
21NIBP-1020 Blood Pressure RANGE 0-500 mmHg @ 20°C ACCURACY +/- (1% of Reading + .5 mmHg)
22 P:\Manuals\BCGroup\…\NIBP1000_UM_Rev06.doc Respiration RATE ACCURACY +/- 1% IMPEDANCE DE
23 NOTES
1 DESCRIPTION... 3 OVERVIEW...
2 Manual NIBP-1000 Series Copyright © 2006 www.bcgroupintl.com Made in the USA 7/06
3 The Model NIBP-1000 Series is a family of Microprocessor based, High Precision Non-Invasive Blood Pressure (NIBP) Simulators. The units are smal
4AVAILABLE MODELS: NIBP1010 BASIC UNIT NIBP1010-P + PEAK NIBP1010-BP + BATTERY & PEAK NIBP1020 BASIC UNIT WITH EKG NIBP1020-PA
5 This section looks at the layout of the NIBP-1010 and gives descriptions of the elements that are present. OVERVIEW 7 pin Mini-Din connector fo
6This section looks at the layout of the NIBP-1020 and gives descriptions of the elements that are present. 7 pin Mini-Din connector for RS-232 and
7 Six tactile-touch keys are provided for system operation: – This key turns the unit off and on. The unit will return to the screen that was a