PRIMING OVERVIEW
PRIMING OVERVIEW
PRIMINGPRIMING
PRIMINGPRIMING
PRIMING
This section covers how to prime the unit prior to running a
Flow, Flow--PCA or Distal Occlusion test using the Infusion
Analyzer. Only Distilled Water can be used with this unit. Do
not use tap water, glucose or any other fluid; this will cause
the interior tubing to become unclean.
For Maximum Accuracy, all air must be cleared from the unit.
This is best done by a high flow rate which is easily obtained
with firm pressure on the priming syringe.